This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of tricyclic antidepressants and metabolites.
The compounds used in this study are –
Compounds |
USP Tailing Factors |
---|---|
Nordoxepin HCL |
1.5 |
Doxepin HCL |
1.5 |
Nortriptyline HCL |
1.3 |
Amtriptyline HCL |
1.2 |
Column: |
Symmetry C18, 3.9 x 150 mm, 5 μm |
Part numbers: |
WAT046980 |
Mobile phase: |
20 mM potassium phosphate, pH 7.0/methanol 29:71 |
Flow rate: |
1.0 mL/min |
Injection volume: |
5 μL of 1 to 1.16 mg/mL |
Temperature: |
35 °C |
Detection: |
UV @ 254 nm |
WA31763.174, June 2003