This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstates analysis of food sugars in milk.
Compounds analysed in this application include:
1. p-Toluamide
2. Fructose
3. Glucose
4. Sucrose
5. Maltose
6. Lactose
Column: |
ACQUITY UPLC BEH Amide 2.1 x 100 mm, 1.7 μm |
Part Number: |
186004801 |
Mobile Phase A: |
80/20 MeCN/H2O with 0.2% triethylamine [TEA] |
Mobile Phase B: |
30/70 MeCN/H2O with 0.2% triethylamine [TEA] |
Flow Rate: |
0.13 mL/min |
Flow Profile: |
90% A/10% B (75% MeCN with 0.2% TEA) |
Injection Volume: |
1.3 μL (PLNO) |
Sample Concentration: |
Standards at 1 mg/mL each |
Sample Diluent: |
50/50 MeCN/H2O |
Column Temperature: |
35°C |
Strong Needle Wash: |
20/80 MeCN/H2O (800 μL) |
Weak Needle Wash: |
75/25 MeCN/H2O (500 μL) |
Seal Wash: |
50/50 MeCN/H2O |
Instrument: |
Waters ACQUITY UPLC with ELSD |
Gain: |
200 |
Pressure: |
40 psi |
Drift Tube Temperature: |
40 °C |
Nebulizer: |
Cooling |
Data Rate: |
10 pps |
Filter Time Constant: |
Normal |
WA60118, September 2009