This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of basic drugs in cell culture lysate using Oasis HLB on-line protocol.
Basic drugs has been analyzed in cell culture lysate using LC-MS/MS in this application brief.
            
        | 
                     Column:  | 
                     XTerra MS C18, 2.1 x 30 mm, 3.5 μm (p/n: 186000398)  | 
            
| 
                     Mobile phase A:  | 
                     Acetonitrile + 0.5% Formic Acid  | 
            
| 
                     Mobile phase B:  | 
                     Water + 0.5% Formic Acid  | 
            
| 
                     Flow rate:  | 
                     1.0 mL  | 
            
| 
                     LC1:  | 
                     Alliance 2690–0.4 mL/min  | 
            
| 
                     LC2:  | 
                     Waters 515–4.0 mL/min  | 
            
| 
                     Loading mobile phase:  | 
                     100 % water  | 
            
| 
                     Eluting mobile phase:  | 
                     1 minute gradient 5% ACN to 95% ACN  | 
            
| 
                     Eluting mobile Phase additive:  | 
                     0.5 % Formic acid  | 
            
| 
                     Extraction column temp.:  | 
                     40 °C  | 
            
| 
                     Switching valve:  | 
                     Rheodyne LabPro 10 ports, 2 position  | 
            
| 
                     MS:  | 
                     Waters Micromass Quattro Ultima  | 
            
| 
                     Source:  | 
                     Electrospray positive  | 
            
| 
                     Sorce temp.:  | 
                     150 °C  | 
            
| 
                     Desolvation gas:  | 
                     600 L/hr  | 
            
| 
                     Gas cell:  | 
                     1.5e-3 mbar  | 
            
| 
                     Cone voltage:  | 
                     20 volts  | 
            
| 
                     Collision energy:  | 
                     20  | 
            
            
        Compound 2 name: Propranolol
Coefficient of Determination: 0.998650
Calibration curve: -1.52393e-5 * x2 + 0.00390583 * x + 0.000102086
Response type: Internal Std (Ref 1), Area * (IS Conc. / IS Area)
Curve type: 2nd Order, Origin: Exclude, Weighting: 1/x2, Axis trans: None
Compound 3 name: Naloxone
Coefficient of Determination: 0.995263
Calibration curve: -2.47393e-5 * x2 + 0.00892751 * x + -5.00665e-5
Response type: Internal Std (Ref 1), Area * (IS Conc. / IS Area)
Curve type: 2nd Order, Origin: Exclude, Weighting: 1/x2, Axis trans: None
            
        Oasis HLB 2.1 x 20 mm, 25 μm Extraction Column Part Number 186000706
            
        
            
        
            
        
            
        
            
        WA31764.112, June 2003