This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates anaylsis of naltrexone in plasma by LC-MS.
The compound analyzed in this study is naltrexone.
Column: |
Symmetry C8, 2.1 x 30 mm, 5 μm |
Part number: |
WAT058977 |
Mobile phase: |
10 mM NH4OH, pH 5/Acetonitrile 88:12 |
Flow rate: |
200 μL/min |
Injection volume: |
50 μL plasma extract |
MS: |
Micromass Quattro II |
Ion source temp.: |
120 °C |
CID: |
1.8e-3 mBar, 20eV |
Desolvation temp.: |
400 °C |
Nebulizer: |
40 L/hr |
Drying gas: |
400 L/hr |
Cone volt: |
56V |
Naltrexone m/z 342 > 324
Naloxone m/z 328 > 310
Coefficient of Determination: 0.996457
Calibration Curve: 0.0508208* x + 0.00306776
Response type: Internal Std (20 ng/mL Naloxone),
Area * ( IS Conc./IS Area)
Curve type: Linear, Origin: Exclude, Weighting: 1x, Axis trans: None
WA31763.108, June 2003