• Note d'application

Prednisolone

Prednisolone

  • Waters Corporation
Ensure consistent quality and safety for your pharmaceutical and biopharmaceutical products with Waters quality control and release testing solutions.

This is an Application Brief and does not contain a detailed Experimental section.

Abstract

This application brief highlights the analysis of prednisolone using XTerra RP18 columns.

Introduction

Prednisolone is studied in this application brief.

Experimental

Conditions

Column:

XTerra RP18 4.6 x 150 mm, 5 μm (p/n: 186000492)

Mobile phase:

H2O/ACN/100 mM NH4COOH, pH 3.0, 35: 55: 10

H2O/ACN/100 mM NH4HCO3, pH 7.0, 35: 55: 10

H2O/ACN/100 mM NH4HCO3, pH 10.0, 58: 32: 10

Flow rate:

1.0 mL/min

Injection volume:

5 μL of 250 μg/mL

Temperature:

30°C

Detection:

UV @ 270 nm

Instrument:

Alliance 2695, 2996 PDA

Mobile phase pH

USP Tailing

3.0

1.19

7.0

1.21

10.0

1.12

 

Results and Discussion

Featured Products

WA20738.090, June 2002

Haut de la page Haut de la page