ACQUITY™ UPLC™ I-Class/Xevo™ TQ Absolute IVD System: Analytical Performance for Immunosuppressive Agents
This is an Application Brief and does not contain a detailed Experimental section.
For in vitro diagnostic use. Not available in all countries.
Abstract
The Waters™ ACQUITY™ UPLC™ I-Class/Xevo™ TQ Absolute IVD System enables the quantification of organic compounds in human biological liquid matrices.
This document describes a test of the analytical performance of the ACQUITY UPLC I-Class/Xevo TQ Absolute IVD System for the analysis of cyclosporine, everolimus, sirolimus, and tacrolimus in whole blood.
Experimental
The ACQUITY UPLC I-Class/Xevo TQ Absolute IVD System was controlled by MassLynx™ IVD (v4.2) and the data processed using the TargetLynx™ Application Manager. Whole blood Calibrators and Quality Controls were processed using the following conditions:
Sample Preparation Conditions
50 µL sample was processed with zinc sulfate, acetonitrile, and centrifuged.
LC Conditions
Column: |
ACQUITY UPLC HSS C18 SB 1.8 µm, 2.1 mm x 30 mm, |
Mobile phase A: |
2 mM Ammonium acetate + 0.1% formic acid in water |
Mobile phase B: |
2 mM Ammonium acetate + 0.1% formic acid in methanol |
Flow rate: |
0.4mL/min |
Gradient: |
50% B for 0.2 minutes, 50–100% B over 0.4 minutes, hold 100% B for 0.6 minutes, equilibrate with 50% B for 0.6 minutes at 0.6 mL/min |
MS Conditions
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75 FWHM) |
Acquisition mode: |
MRM |
Polarity: |
ESI+ |
Results and Discussion
Performance characteristics of cyclosporine, everolimus, sirolimus, and tacrolimus on the ACQUITY UPLC I-Class/Xevo TQ Absolute IVD System are shown in Table 1. Analytical sensitivity of the chromatographic separation is illustrated in Figure 1.


Conclusion
The Waters ACQUITY UPLC I-Class/Xevo TQ Absolute IVD System has demonstrated the capability to deliver analytical sensitivity and precision for the analysis of cyclosporine, everolimus, sirolimus, and tacrolimus in whole blood.
Disclaimer
The analytical performance data presented here is for illustrative purposes only. Waters does not recommend or suggest analysis of the analytes described herein. These data are intended solely to demonstrate the performance capabilities of the system for analytes representative of those commonly analyzed using liquid chromatography and tandem mass spectrometry. Performance in an individual laboratory may differ due to a number of factors, including laboratory methods, materials used, intra-operator technique, and system conditions. This document does not constitute a warranty of merchantability or fitness for any particular purpose, express, or implied, including for the testing of the analytes in this analysis.
Featured Products
720007886, March 2023