This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates the LC-MS gradient separation of buprenorphine and buprenorphine glucuronide on XBridge HILIC Columns.
The compounds used in this study are:
Column: |
XBridge HILIC, 3.5 μm, 2.1 x 100 mm |
Part number: |
186004433 |
Mobile phase A: |
10 mM NH4COOH in H2O, 0.125% HCOOH in 50:50 ACN:H2O |
Mobile phase B: |
10 mM NH4COOH in H2O, 0.125% HCOOH in 90:10 ACN:H2O |
Flow rate: |
0.3 mL/min |
Injection volume: |
5 μL |
Sample concentration: |
50 ng/mL each |
Sample diluent: |
75:25 ACN:MeOH with 0.2% HCOOH |
Column temperature: |
30 °C |
Instrument: |
Waters ACQUITY UPLC with TQD |
MRM: |
Buprenorphine 468.3 > 54.7 Buprenorphine glucuronide 644.3 > 468.3 |
Dwell time: |
75 ms |
ISD: |
10 ms |
ICD: |
10 ms |
Time(min) |
Profile |
|
---|---|---|
%A |
%B |
|
0.00 |
0.1 |
99.9 |
1.05 |
0.1 |
99.9 |
4.35 |
99.9 |
0.1 |
5.00 |
99.9 |
0.1 |
5.01 |
0.1 |
99.9 |
6.00 |
0.1 |
99.9 |
WA64077, August 2009