This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of niflumic acid/camphor sulfonic acid extracted from rat plasma.
The compounds used in this study are - Niflumic Acid and (1S)-Camphor-10-sulfonic acid.
            
        | 
                     Plate:  | 
                     Oasis WAX μElution Plate  | 
            
| 
                     Condition:  | 
                     200 μL MeOH  | 
            
| 
                     Equilibrate:  | 
                     200 μL H2O  | 
            
| 
                     Load:  | 
                     100 μL 1:1 diluted rat plasma, 10 pg/μL, w/2% H3PO4  | 
            
| 
                     Wash 1:  | 
                     200 μL 25 mM NH4CH2COOH buffer in H2O, pH 5  | 
            
| 
                     Wash 2:  | 
                     200 μL MeOH  | 
            
| 
                     Elute:  | 
                     50 μL (25 μL x 2) MeOH w/2% NH4OH  | 
            
| 
                     Dilute:  | 
                     50 μL H2O with 2% HCOOH  | 
            
| 
                     Inject:  | 
                     10 μL  | 
            
| 
                     Column:  | 
                     SunFire C18 2.1 x 20 mm IS, 3.5 μm  | 
            ||
| 
                     Part number:  | 
                     186002531  | 
            ||
| 
                     Mobile phase A:  | 
                     H2O with 10 mM NH4CH3COOH, pH 5.5  | 
||
| 
                     Mobile phase B:  | 
                     MeOH with 10 mM NH4CH3COOH, pH 5.5  | 
||
| 
                     Flow rate:  | 
                     0.4 mL/min  | 
            ||
| 
                     Injection volume:  | 
                     10 μL  | 
            ||
| 
                     Sample concentration:  | 
                     10 pg/μL each analyte  | 
            ||
| 
                     Temperature:  | 
                     Ambient  | 
            ||
| 
                     Instrument:  | 
                     Waters Alliance 2695 Waters Micromass Quattro micro API  | 
            ||
| 
                     ESI source temp.:  | 
                     150 °C  | 
            ||
| 
                     Desolvation temp.:  | 
                     350 °C  | 
            ||
| 
                     Cone gas flow:  | 
                     50 L/Hr  | 
            ||
| 
                     Desolvation gas flow:  | 
                     600 L/Hr  | 
            ||
| 
                     Collision Cell:  | 
                     1.23e-3 bar (Ar gas)  | 
            ||
| 
                     Compounds  | 
                     MRM Transition  | 
                     Cone (V)  | 
                     CID(eV)  | 
            
| 
                     Niflumic Acid  | 
                     m/z 281.1 → 237.1  | 
                     30  | 
                     17  | 
            
| 
                     Camphorsulfonic Acid  | 
                     m/z 231.1 → 79.9  | 
                     45  | 
                     30  | 
            
            
        
            
        
            
        WA41895, May 2005