This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes the gradient separation of catecholamines on Atlantis HILIC Silica column.
The compounds used in this study are:
Column: |
Atlantis HILIC Silica, 4.6 x 50 mm, 5 μm |
Part Number: |
186002028 |
Mobile Phase A: |
H2O |
Mobile Phase B: |
ACN |
Mobile Phase C: |
200 mM NH4COOH, pH 3.0 |
Flow Rate: |
2.0 mL/min |
Injection Volume: |
5.0 μL |
Sample Diluent: |
75:25 ACN:MeOH with 0.2% FA |
Sample Concentration: |
42 μg/mL Norepinephrine (NE), 17 μg/mL Epinephrine (E), 83 μg/mL 3-hydroxytyramine (Dopamine) |
Detection: |
UV @ 280 nm |
Temperature: |
Ambient |
Instrument: |
Waters Alliance System, 2695 Separations Module with 2996 PDA |
Time(min) |
Profile |
||
---|---|---|---|
%A |
%B |
%C |
|
0.0 |
0 |
95 |
5 |
5.0 |
25 |
7 |
5 |
WA64073, August 2009