This application describes methods to determine malachite green in fish using UPLC-MS/MS.

Introduction

Malachite Green (MG) is an effective and inexpensive fungicide used in aquaculture. During metabolism MG reduces to Leucomalachite Green (LMG), which has been shown to accumlate in fatty fish tissues. Both MG and LMG have demonstrated putative carcinogenic activity, and thus, have been banned for use in aquaculture by both the United States Food and Drug Administration (US FDA) and European Union (EU).

Experimental

Pre-treatment

1. Weigh 1 g sample into a 30 mL centrifuge tube.

2. Add 50 μL TMPD* solution (1 mg/mL).

3. Add standard solutions MG, LMG, 0.1 μg/mL and internal standard, leave to stand for 10 minutes.

4. Add 10 mL McIlvaines Buffer (pH 2.6) /methanol (50:50 v/v) solution; homogenize for 45 seconds.

5. Centrifuge at 5000 rpm for 20 minutes, transfer supernatant into centrifuge tube.

6. Repeat steps 4 and 5, combine the two portions of supernatant. A 20 mL aliquot will be used for SPE.

*TMPD= N, N, N’, M’- Tetramethyl-1,4-phenylenediamine dihydrochloride

SPE Procedure

Solutions

McIlvaines Buffer (pH 2.6) :

1. 0.1 M citric acid monohydrate (A) - Dissolve citric acid monohydrate (10.5 g) in water (500 mL).

2. 0.2 M disodium hydrogen phosphate dihydrate (B) - Dissolve disodium hydrogen phosphate dihydrate (14.2 g) in water (500 mL).

3. Mix A (445.5 mL) and B (54.5 mL).

McIlvaines Buffer (pH 2.6): methanol (50:50 v/v):

Mix McIlvaines Buffer (pH 2.6) (250 mL) with methanol (250 mL).

LC Conditions

System:

ACQUITY UPLC

Column:

ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 50 mm

Flow rate:

0.4 mL/min

Mobile phase A:

0.1% formic acid in water

Mobile phase B:

0.1% formic acid in methanol

Gradient

MS Conditions

MS System:

Waters Quattro Premier

Ionization mode:

Positive electrospray (ESI+)

Multiple reaction monitoring

MRM method parameters
MRM method parameters. 

Results and Discussion

The LOD of LMG and MG
The LOD of LMG and MG are 0.02 ppb and 0.01 ppb, respectively. The recoveries of LMG and MG in fish is between 50–80%.  