ACQUITY UPLC I-Class/Xevo TQD IVD System: Analytical Performance for an Inosine-5'-Monophosphate Dehydrogenase Inhibitor
Este é um Resumo de aplicações e, por isso, não inclui uma seção de experimento detalhada.
Para uso em diagnóstico in vitro. Não está disponível para venda em todos os países.
Introduction
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System enables the quantification of organic compounds in human biological liquid matrices.
This document describes a test of the analytical performance of the ACQUITY UPLC I-Class/Xevo TQD IVD System for the analysis of mycophenolic acid in plasma.

Experimental
The ACQUITY UPLC I-Class/Xevo TQD IVD System was controlled by MassLynx IVD Software (v4.1) and the data processed using the TargetLynx Application Manager. Commercial calibrators and Quality Controls were used as well as in-house calibrators prepared by spiking commercially available reference material in plasma. The samples were processed using the following conditions:
Sample Preparation Conditions
50 μL sample was processed with zinc sulphate/methanol and centrifuged. The supernatants were transferred to a collection plate for analysis.
LC Conditions
Column: |
ACQUITY UPLC HSS C18 SB 1.8 μm, 2.1 mm × 30 mm |
Mobile phase A: |
2 mM Ammonium acetate+0.1% formic acid in water |
Mobile phase B: |
2 mM Ammonium acetate+0.1% formic acid in methanol |
Flow rate: |
0.7 mL/min |
Gradient: |
30–40% B over 0.75 minutes, 40–75% B over 0.85 minutes, 98% B for 0.4 minutes |
MS Conditions
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75 FWHM) |
Acquisition mode: |
MRM |
Polarity: |
ESI (+) |
Results and Discussion
Performance characteristics of mycophenolic acid on the ACQUITY UPLC I-Class/Xevo TQD IVD System are shown in Table 1. A chromatogram illustrating the selectivity of the mycophenolic acid analysis is shown in Figure 1.


Conclusion
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System has demonstrated the capability to deliver analytically sensitive, accurate, and precise performance for mycophenolic acid in plasma.
Disclaimer
The analytical performance data presented here is for illustrative purposes only. Waters does not recommend or suggest analysis of the analytes described herein. These data are intended solely to demonstrate the performance capabilities of the system for analytes representative of those commonly analyzed using liquid chromatography and tandem mass spectrometry. Performance in an individual laboratory may differ due to a number of factors, including laboratory methods, materials used, intra-operator technique, and system conditions. This document does not constitute a warranty of merchantability or fitness for any particular purpose, express or implied, including for the testing of the analytes in this analysis.
720006340, August 2018