ACQUITY™ UPLC™ I-Class/Xevo™ TQD IVD System: Analytical Performance for Antidepressant Drugs
Ce document est une note d’application et ne contient pas de section détaillée concernant l’expérimentation.
Pour usage diagnostique in vitro. La disponibilité varie selon les pays.
Abstract
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System enables the quantification of organic compounds in human biological liquid matrices.
This document describes a test of the analytical performance of the ACQUITY UPLC I-Class/Xevo TQD IVD System for the analysis of citalopram, desmethylfluoxetine, duloxetine, fluoxetine, fluvoxamine, O-desmethylvenlafaxine, mirtazapine, sertraline, trazodone, and venlafaxine in plasma.
Experimental
The ACQUITY UPLC I-Class/Xevo TQD IVD System was controlled by MassLynx™ Software (v4.2) and the data processed using the TargetLynx™ XS Application Manager. Calibrators and Quality Controls were prepared by spiking commercially available reference material in plasma and the samples were processed using the following conditions:
Sample Preparation Conditions
A 50 µL sample was processed with acetonitrile and centrifuged, then subsequently diluted with water prior to analysis.
LC Conditions
Column: |
XSelect™ Premier HSS T3, 2.5 µm, 2.1 mm x 100 mm |
|
Mobile phase A: |
2 mM ammonium acetate in water |
|
Mobile phase B: |
2 mM ammonium acetate in methanol |
|
Flow rate: |
0.4 mL/min / 0.6 mL/min |
|
Gradient: |
40% A initial, hold until 1.70 minutes, then ramp gradient 6 until 20% A at 4.00 minutes, at 4.01 minutes switch to 100% B at a flow rate of 0.6 mL/min until 4.50 minutes, finally return to 40% A at 0.4 mL/min from 4.50 until 5.0 minutes. |
|
MS Conditions
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75FWHM) |
Acquisition mode: |
MRM |
Polarity: |
ESI+ |
Results and Discussion
Chromatographic selectivity of a range of antidepressants using the ACQUITY UPLC I-Class/Xevo TQD IVD System is illustrated in Figure 1. Performance characteristics of the antidepressant drugs are shown in Table 1.
Conclusion
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System has demonstrated the capability to analyze a panel of antidepressant drugs in plasma.
Disclaimer
The analytical performance data presented here is for illustrative purposes only. Waters does not recommend or suggest analysis of the analytes described herein. These data are intended solely to demonstrate the performance capabilities of the system for analytes representative of those commonly analyzed using liquid chromatography and tandem mass spectrometry. Performance in an individual laboratory may differ due to a number of factors, including laboratory methods, materials used, intra-operator technique, and system conditions. This document does not constitute a warranty of merchantability or fitness for any particular purpose, express or implied, including for the testing of the analytes in this analysis.
720007860, February 2023