This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of catecholamines using XBridge Shield RP18 Columns.
Catecholamines, such as epinephrine (adrenaline), norepinephrine (noradrenaline), and dopamine, are derived from the amino acid tyrosine and act as important hormones or neurotransmitters.
Columns: |
XBridge Shield RP18 4.6 X 100 mm, 3.5 μm p/n: 186003044 |
Mobile phase A: |
H2O |
Mobile phase B: |
ACN |
Mobile phase C: |
100 mM HCOONH4, pH 3.0 |
Flow rate: |
1.0 mL/min |
Injection volume: |
10 μL |
Sample Concentration and Diluent: |
10 μg/mL in H2O |
Temp.: |
30 °C |
Sampling Rate: |
5 points /second |
Detection: |
UV @ 280 nm |
Time Constant: |
1.0 |
Needle Wash: |
5/95 MeOH/H2O |
Instrument: |
Alliance 2695 with 2996 PDA |
Time(min) |
%A |
%B |
%C |
---|---|---|---|
0 |
90 |
0 |
10 |
5 |
90 |
0 |
10 |
15 |
65 |
25 |
10 |
16 |
65 |
25 |
10 |
17 |
90 |
0 |
10 |
20 |
90 |
0 |
10 |
WA60201, June 2007