• Application Note

Tetracyclines in Serum

Tetracyclines in Serum

  • Waters Corporation
Plasma serum preparation, vials in tray

This is an Application Brief and does not contain a detailed Experimental section.

Abstract

This application brief highlights the analysis of tetracyclines in serum using Symmetry Shield RP8 Columns. 

Introduction

The compounds analyzed in this study are: 

  1. Minocycline 
  2. Tetracycline 
  3. Demeclocycline (I.S.) 
Compounds

Experimental

HPLC Method

Column:

Symmetry Shield RP8, 3.0 x 150 mm, 5 μm

Guard column:

Symmetry guard column, 3.9 x 20 mm, 5μm

Part numbers:

Column - WAT094243, Guard - WAT200675

Mobile phase:

0.1% TFA in Water, Methanol, Acetonitrile 91:2:7

Flow rate:

0.90 mL/min

Injection volume:

20 μL of reconstituted porcine serum extract

Detection:

UV @ 350 nm

Extraction Method

Results and Discussion

% Recovery
Chromatogram of Serum Extracts: A) Blank B) Spiked Sample
Chromatogram of Serum Extracts: A) Blank B) Spiked Sample. 

Featured Products

WA31763.168, June 2003