        This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates anaylsis of naltrexone in plasma by LC-MS.
The compound analyzed in this study is naltrexone.
| Column: | Symmetry C8, 2.1 x 30 mm, 5 μm | 
| Part number: | WAT058977 | 
| Mobile phase: | 10 mM NH4OH, pH 5/Acetonitrile 88:12 | 
| Flow rate: | 200 μL/min | 
| Injection volume: | 50 μL plasma extract | 
| MS: | Micromass Quattro II | 
| Ion source temp.: | 120 °C | 
| CID: | 1.8e-3 mBar, 20eV | 
| Desolvation temp.: | 400 °C | 
| Nebulizer: | 40 L/hr | 
| Drying gas: | 400 L/hr | 
| Cone volt: | 56V | 
 
            
         
            
        Naltrexone m/z 342 > 324
Naloxone m/z 328 > 310
 
            
         
            
         
            
        Coefficient of Determination: 0.996457
Calibration Curve: 0.0508208* x + 0.00306776
Response type: Internal Std (20 ng/mL Naloxone),
Area * ( IS Conc./IS Area)
Curve type: Linear, Origin: Exclude, Weighting: 1x, Axis trans: None
WA31763.108, June 2003