This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights on the analysis of codeine and its glucuronide metabolite in human urine.
The compounds analyzed in this study are:
            
        
            
        | 
                     Column:  | 
                     Symmetry C18, 2.1 x 150 mm, 3.5 μm  | 
            
| 
                     Guard column:  | 
                     Symmetry C18, 2.1 x 10 mm, 3.5 μm  | 
            
| 
                     Part numbers:  | 
                     Column - 186000174, Guard - 186000169  | 
            
| 
                     Mobile phase:  | 
                     Methanol/Acetonitrile/0.05% TFA in water 5:5:90  | 
            
| 
                     Flow rate:  | 
                     0.3 mL/min  | 
            
| 
                     Injection volume:  | 
                     80 μL urine extract  | 
            
| 
                     Temperature:  | 
                     30° C  | 
            
| 
                     Detection:  | 
                     UV @ 220 nm (0.04 AUFS)  | 
            
            
        
            
        
            
        Test subject took a therapeutic 60 mg oral dose of Codeine and subsequently had urine monitored over a 21 hour timeframe. Peaks of interest were confirmed by LC-MS, at the 10.3 hour timepoint, to be Codeine-6-Glucuronide and Codeine
            
        WA31764.63, June 2003