 
            
        This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes UPLC-MS analysis of food sugars/saccharides in beer.
Compounds analysed in beer are:
 
            
        | Column: | ACQUITY UPLC BEH Amide 2.1 x 100 mm, 1.7 μm | 
| Part Number: | 186004801 | 
| Mobile Phase A: | 80/20 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH] | 
| Mobile Phase B: | 30/70 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH] | 
| Flow Rate: | 0.13 mL/min | 
| Gradient: | 10 minute gradient, 75%-45% MeCN (w/0.1% NH4OH) with 25 minute re-equilibration | 
| Injection Volume: | 2.0 μL (PLNO) | 
| Sample Concentration: | Standards at 10 μg/mL, Beer at 50% dilution | 
| Sample Diluent: | 50/50 MeCN/H2O | 
| Column Temperature: | 35 °C | 
| Strong Needle Wash: | 20/80 MeCN/H2O (800 μL) | 
| Weak Needle Wash: | 75/2MeCN/H2O 5 (500 μL) | 
| Seal Wash: | 50/50 MeCN/H2O | 
| Instrument: | Waters ACQUITY UPLC with SQ | 
| Time(min) | Profile | |
|---|---|---|
| %A | %B | |
| 0.00 | 90.00 | 10.00 | 
| 10.00 | 30.00 | 70.00 | 
| 10.01 | 90.00 | 10.00 | 
| 35.00 | 90.00 | 10.00 | 
| Ionization Mode: | ES- | 
| Capillary: | 2.8 kV | 
| Cone Voltage: | 25 V (fructose, glucose, maltotriose); 40V (sucrose and maltose) | 
| Source Temp: | 120 °C | 
| Desolvation Temp: | 350 °C | 
| Desolvation Gas Flow: | 500 L/Hr | 
| Cone: | 50 L/Hr | 
| SIR (m/z): | 179.0 (fructose, glucose); 341.1 (sucrose, maltose); 503.2 (maltotriose) | 
| Dwell Time: | 0.04 s | 
 
            
        WA60126, October 2009