This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes the analysis of food sugars in prepared food using ACQUITY UPLC BEH Amide Columns.
Compounds used for this study includes:
1. p-Toluamide
2. Fructose
3. Glucose
4. Sucrose
5. Maltose
6. Lactose
            
        | 
                     Column:  | 
                     ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm  | 
            
| 
                     Part number:  | 
                     186004800  | 
            
| 
                     Mobile phase A:  | 
                     80/20 acetone/H2O with 0.05% triethylamine [TEA]  | 
            
| 
                     Mobile phase B:  | 
                     30/70 acetone/H2O with 0.05% triethylamine [TEA]  | 
            
| 
                     Flow rate:  | 
                     0.15 mL/min  | 
            
| 
                     Flow profile:  | 
                     95% A/5% B (77.5% acetone with 0.05% TEA)  | 
            
| 
                     Injection volume:  | 
                     0.7 μL (PLNO)  | 
            
| 
                     Sample concentration:  | 
                     Standards at 1 mg/mL each  | 
            
| 
                     Sample diluent:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Column temperature:  | 
                     85 °C  | 
            
| 
                     Strong needle wash:  | 
                     20/80 MeCN/H2O (800 μL)  | 
            
| 
                     Weak needle wash:  | 
                     75/25 MeCN/H2O (500 μL)  | 
            
| 
                     Seal wash:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Instrument:  | 
                     Waters ACQUITY UPLC with ELSD  | 
            
| 
                     Gain:  | 
                     200  | 
            
| 
                     Pressure:  | 
                     40 psi  | 
            
| 
                     Drift tube temperature:  | 
                     40 °C  | 
            
| 
                     Nebulizer:  | 
                     Cooling  | 
            
| 
                     Data rate:  | 
                     10 pps  | 
            
| 
                     Filter time constant:  | 
                     Normal  | 
            
            
        WA60115, October 2009