ACQUITY UPLC I-Class/Xevo TQD IVD System: Analytical Performance for Fat Soluble Vitamins
Dies ist ein Applikationsbericht, der keinen detaillierten Abschnitt zu Versuchen enthält.
Zur Verwendung in der In-vitro-Diagnostik. Nicht in allen Ländern erhältlich.
Introduction
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System enables the quantification of organic compounds in human biological liquid matrices.
This document describes a test of the analytical performance of the ACQUITY UPLC I-Class/Xevo TQD IVD System for the analysis of vitamin A and vitamin E in serum.

Experimental
The ACQUITY UPLC I-Class/Xevo TQD IVD System was controlled by MassLynx IVD Software (v4.2) and the data was processed using the TargetLynx Application Manager. In-house calibrators and quality controls were prepared by spiking commercially available reference material in serum. The samples were processed using the following conditions:
Sample Preparation Conditions
A 100-μL sample was processed with ethanol and water and centrifuged. The supernatants were diluted then loaded onto Oasis PRiME HLB μElution plates, washed, and eluted prior to analysis.
LC Conditions
Column: |
ACQUITY UPLC HSS PFP, 1.8 μm, 2.1 mm × 50 mm |
Mobile phase A: |
2 mM ammonium acetate + 0.1% formic acid in water |
Mobile phase B: |
2 mM ammonium acetate + 0.1% formic acid in methanol |
Flow rate: |
0.4 mL/min |
Gradient: |
65–98% B over 2 minutes, 98% B for 0.55 minutes, 65% A for 0.95 minutes |
MS Conditions
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75 FWHM) |
Acquisition mode: |
MRM |
Polarity: |
ESI+ |
Results and Discussion
Chromatograms illustrating the chromatographic selectivity of vitamins A and E are shown in Figures 1 and 2. Performance characteristics of vitamins A and E on the ACQUITY UPLC I-Class/Xevo TQD IVD System are shown in Table 1.



Conclusion
The Waters ACQUITY UPLC I-Class/Xevo TQD IVD System has demonstrated the capability to deliver analytically sensitive, accurate, and precise performance for vitamins A and E in serum.
Disclaimer
The analytical performance data presented here is for illustrative purposes only. Waters does not recommend or suggest analysis of the analytes described herein. These data are intended solely to demonstrate the performance capabilities of the system for analytes representative of those commonly analyzed using liquid chromatography and tandem mass spectrometry. Performance in an individual laboratory may differ due to a number of factors, including laboratory methods, materials used, intra-operator technique, and system conditions. This document does not constitute a warranty of merchantability or fitness for any particular purpose, express or implied, including for the testing of the analytes in this analysis.
720006728, December 2019