This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes analysis of food sugars in sport drink using ACQUITY UPLC BEH Amide Column.
Compounds used for this study includes:
1. p-Toluamide
2. Fructose
3. Glucose
4. Sucrose
5. Maltose
6. Lactose
 
            
        | Column: | ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm | 
| Part number: | 186004800 | 
| Mobile phase A: | 80/20 acetone/H2O with 0.05% triethylamine [TEA] | 
| Mobile phase B: | 30/70 acetone/H2O with 0.05% triethylamine [TEA] | 
| Flow rate: | 0.15 mL/min | 
| Flow profile: | 95% A/5% B (77.5% acetone with 0.05% TEA) | 
| Injection volume: | 0.7 μL (PLNO) | 
| Sample concentration: | Standards at 1 mg/mL each | 
| Sample diluent: | 50/50 MeCN/H2O | 
| Column temperature: | 85 °C | 
| Strong needle wash: | 20/80 MeCN/H2O (800 μL) | 
| Weak needle wash: | 75/25 MeCN/H2O (500 μL) | 
| Seal wash: | 50/50 MeCN/H2O | 
| Instrument: | Waters ACQUITY UPLC with ELSD | 
| Gain: | 200 | 
| Pressure: | 40 psi | 
| Drift tube temperature: | 40 °C | 
| Nebulizer: | Cooling | 
| Data rate: | 10 pps | 
| Filter time constant: | Normal | 
 
            
        WA60116, October 2009