This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of niflumic acid/camphor sulfonic acid extracted from rat plasma.
The compounds used in this study are - Niflumic Acid and (1S)-Camphor-10-sulfonic acid.
Plate: |
Oasis WAX μElution Plate |
Condition: |
200 μL MeOH |
Equilibrate: |
200 μL H2O |
Load: |
100 μL 1:1 diluted rat plasma, 10 pg/μL, w/2% H3PO4 |
Wash 1: |
200 μL 25 mM NH4CH2COOH buffer in H2O, pH 5 |
Wash 2: |
200 μL MeOH |
Elute: |
50 μL (25 μL x 2) MeOH w/2% NH4OH |
Dilute: |
50 μL H2O with 2% HCOOH |
Inject: |
10 μL |
Column: |
SunFire C18 2.1 x 20 mm IS, 3.5 μm |
||
Part number: |
186002531 |
||
Mobile phase A: |
H2O with 10 mM NH4CH3COOH, pH 5.5 |
||
Mobile phase B: |
MeOH with 10 mM NH4CH3COOH, pH 5.5 |
||
Flow rate: |
0.4 mL/min |
||
Injection volume: |
10 μL |
||
Sample concentration: |
10 pg/μL each analyte |
||
Temperature: |
Ambient |
||
Instrument: |
Waters Alliance 2695 Waters Micromass Quattro micro API |
||
ESI source temp.: |
150 °C |
||
Desolvation temp.: |
350 °C |
||
Cone gas flow: |
50 L/Hr |
||
Desolvation gas flow: |
600 L/Hr |
||
Collision Cell: |
1.23e-3 bar (Ar gas) |
||
Compounds |
MRM Transition |
Cone (V) |
CID(eV) |
Niflumic Acid |
m/z 281.1 → 237.1 |
30 |
17 |
Camphorsulfonic Acid |
m/z 231.1 → 79.9 |
45 |
30 |
WA41895, May 2005