Improved Yield and Productivity of Ionizable Compounds Employing Hybrid Packings.

Library Number:
WA20726
Part Number:
WA20726
Author(s):
C. Mazza, J. Mazzeo, T. Sirard, K. Tran, K. Zheng, Z. Lu, U. Neue [Waters]
Content Type:
Posters
Content Subtype:
Other Symposium
Compounds:
Doxylamine; Diphenhydramine; Oxybutinin; Terfenadine; Propyl Gallate; Oxacillin; Cloxacillin; Dicloxacillin
Column:
XTerra Prep MS C18 5 µm Steel 19 mm 50 mm
Related Products:
 
Sample loading onto a column can be highly compromised due to the solubility of the compounds or their compatibility with the initial mobile phase conditions, resulting in low yields and productivity. Consequently, exploring alternative ways to load sample onto the column is urgently needed. The at-column-dilution method is a significant alternative as it allows an increase in sample loading, improving peak shapes, providing higher yield and productivity of the targeted compound. Case studies are shown where the benefits of the at-column-dilution method are demonstrated when loading basic compounds under acidic conditions and eluting at high pH, as well as when loading samples dissolved in strong sample solvents such as DMSO. The cases shown here illustrate the utility of the at-column-dilution technique while maintaining present day isolation and purification needs both in combinatorial chemistry as well as drug development.

Title Format File Size
wa20726 PDF 2041.3kB