Reliable and Robust Injector Precision and Linearity with the Arc HPLC System

Library Number:
APNT135077577
Part Number:
720007082EN
Author(s):
B. Tirupateswara Rao, Bheeshmacharyulu. S, Dilshad Pullancheri, Dr. Padmakar Wagh
Source:
Waters
Content Type:
Application Notes
Content Subtype:
Application Notes
Related Products:
 
 

Precision and linearity are two measurable characteristics that determine how well an HPLC injector operates under various chromatographic conditions. Analytical methods that have 100% organic sample diluent can have a large impact on the performance of the sample delivery system. The work presented here demonstrates that the new Arc HPLC System delivers linear and highly reproducible sample injections across the range of analytical HPLC injection volumes even when the sample diluent is 100% organic.


Title Format File Size