Improving the Detection of Thyroglobulin in Human Plasma for Clinical Research by Combining SISCAPA Enrichment and Microflow LC-MS

Library Number:
APNT134866663
Part Number:
720005509EN
Author(s):
Jay S. Johnson, Morteza Razavi, James Murphy, Selena Larkin, Paul Rainville
Source:
Waters, SISCAPA Assay Technologies
Content Type:
Application Notes
Content Subtype:
Application Notes
Related Products:
 
 
 
 

This application note investigates if the ionKey/MS System operating in a dual-pump trapping configuration can provide reductions in LLOQ levels for Thyroglobulin (Tg).

The above procedure does provide acceptable LLOQ levels for Tg using significantly less plasma while maintaining the requisite accuracy, better precision, and throughput levels exemplified by standard flow LC-MS methods. Accordingly, microflow is a viable and attractive solution for clinical research.