A simple and rapid analytical method was developed for the determination of seven illegal veterinary drug additives in animal feed. This method has been proven to achieve levels of detection that meet regulatory requirements.
The Oasis PRiME HLB Cartridge provided effective clean-up and good recoveries for the target veterinary drugs in animal feeds.
The ACQUITY UPLC I-Class System coupled with Xevo TQ-S micro offered good sensitivity and robust methodology.
When discussing illegal additives in the feed, we immediately think of clenbuterol. Since the scandal of clenbuterol was exposed in 2011, the Chinese government has established a strict standard for the use of additives, and also tightened regulation for the illegal use of additives in feed. Nevertheless, some feed producers still have not stopped their illegal behavior. The “CCTV 3.15 party in 2017” exposed this situation and aroused great concern from the public. A reporter’s survey found the abuse of veterinary drugs including olaquindox, neomycin sulfate, sulfaquinoxaline, and dihdropyridine in animal breeding.
Olaquindox is an alternative to clenbuterol, that can promote growth, reduce the feed and meat ratio, improve body size, and improve feed intake. It tends to be accumulated in animal tissue and leads to chromosomal abnormalities in cells if added to animal feed over a long period of time. However, the residues of these compounds also pose a health risk to the consumers.
Currently, the determination of olaquindox in the Chinese national standard (GB)1 is mainly based on LC-UV and LC-MS/MS methods. Accurate quantification of Olaquindox is a challenge because of the complex matrices and potential to decompose during sample preparation and when exposed to light.
In this application note, a simple clean-up protocol using a novel SPE device was introduced for the analysis Olaquindox and six other illegal veterinary additives in animal feed. The extract was cleaned up by pass-through SPE using the Oasis PRiME HLB Cartridge prior to UPLC-MS/MS analysis. The spiked samples were quantified using an external standards method, and the recovery and reproducibility for each compound met the regulatory requirements of the quantitative method. This method is simple, rapid, accurate, suitable for the analysis of the highlighted veterinary drugs in animal feed.
LC system: |
ACQUITY UPLC I-Class |
Column: |
ACQUITY UPLC HSS T3, 1.8 μm, 2.1 x 100 mm |
Temp.: |
45 °C |
Flow rate: |
0.4 mL/min |
Mobile phase A: |
0.1% formic acid in water |
Mobile phase B: |
0.1% formic acid in methanol |
Run time: |
9 min |
Injection vol.: |
2 μL |
Time (min) |
Flow rate (mL/min) |
%A |
%B |
---|---|---|---|
0.00 |
0.4 |
98 |
2 |
0.25 |
0.4 |
98 |
2 |
3.25 |
0.4 |
70 |
30 |
7.00 |
0.4 |
2 |
98 |
7.50 |
0.4 |
2 |
98 |
7.60 |
0.4 |
98 |
2 |
9.00 |
0.4 |
98 |
2 |
MS system: |
Xevo TQ-S micro |
Ionization mode: |
ESI+ |
Capillary voltage: |
3.0 kV |
Desolvation temp.: |
550 °C |
Source temp.: |
150 °C |
Desolvation flow: |
1000 L/h |
Cone gas: |
50 L/h |
Step 1: |
Weigh 1 g of feed sample into a 50 mL centrifuge tube; |
Step 2: |
Add 10 mL of extraction solvent (80% acetonitrile + 20% water) and shake well for 10 min; |
Step 3: |
Centrifuge at 6000 rpm for 5 min |
Step 1: |
An Oasis PRiME HLB Cartridge (6 cc, 200 mg; p/n 186008057) was mounted on a pre-cleaned SPE vacuum manifold. Cartridge conditioning is not required and is not performed. |
Step 2: |
A 0.5 mL aliquot of the supernatant (sample extract) was passed through the Oasis PRiME HLB Cartridge and the eluant was discarded. |
Step 3: |
Install the collection tubes. Another 1 mL of supernatant was passed through the cartridge, and the eluant was collected. The eluant was diluted 1:3 with water and injected into Xevo TQ-S micro for analysis. |
The analyte recovery was determined by spiking standards into the blank matrix, a 1:1 mixture of rice and corn powders. The analytes were spiked at concentrations of low, medium (5 times low spike) and high levels (10 times low spike). The lowest spike for olaquindox was 10 ug/kg, sulfaquinoxaline was 0.5 ug/kg, trimethoprim and atropine was 2.5 ug/kg and the lowest spike for aminophylline, diprophylline and dexamethasone was 5.0 ug/kg). Each level of spiking was repeated in five replicates. All samples were processed according to the method described previously. The concentrations were calculated using a matrix-matched calibration curve. The recovery range of the high, medium, and low level samples ranged from 70.6% to 112%. The precision range of the high and medium level spike samples was 0.88% to 4.2% and the precision range was 4.3% to 8.8% for the low spike samples.
The matrix effect was measured by comparing the peak area of solvent standards and post spiked samples in chicken feed and swine feed samples, where the spiked level was equal to 5 µg/kg for atropine and diprophylline, and 1 µg/kg for the other compounds.
Calibration curves ranged from 0.01 to 1.00 µg/L for sulfaquinoxaline, from 0.1 to 10 µg/L for olaquindox, aminophylline, diprophylline, and dexamethasone, and 0.05 to 5.0 µg/L for trimethoprim and atropine.
The established method was used for real sample analysis. Finally, an olaquindox content up to 1.9 to 18 mg/kg was detected in chicken feed and swine feed samples.
720006192, February 2018