This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates the LC-MS isocratic separation of neurotransmitters on Atlantis HILIC Silica Columns.
The compounds used in this study are:
|
Column: |
Atlantis HILIC Silica, 2.1 x 50 mm, 3 μm |
|
Part Number: |
186002011 |
|
Mobile Phase A: |
10 mM NH4COOH with 0.125% HCOOH in H2O |
|
Mobile Phase B: |
10 mM NH4COOH with 0.125% HCOOH in 90:5:5 ACN:MeOH:H2O |
|
Isocratic Mobile Phase Composition: |
10% A; 90% B |
|
Flow Rate: |
0.5 mL/min |
|
Injection Volume: |
10.0 μL (full loop) |
|
Sample Concentration: |
5 ng/mL each |
|
Sample Diluent: |
75:25 ACN:MeOH with 0.2% HCOOH |
|
Column Temp.: |
30 °C |
|
Weak and Strong Needle Wash: |
95;5 ACN:H2O |
|
Detection: |
MS |
|
Sampling Rate: |
5 points/sec |
|
Instrument: |
Waters ACQUITY UPLC with TQD |
|
Ionization Mode: |
ES+ |
|
Capillary: |
0.5 kV |
|
Cone: |
25 V (acetylcholine), 40 V (choline) |
|
Source Temperature: |
120 °C |
|
Desolvation Temperature: |
350 °C |
|
Desolvation Gas Flow: |
800 L/Hr |
|
SIR: |
146.1 m/z (acetylcholine); 104.0 m/z (choline) |
|
Dwell Time: |
150 msec |
|
ISD: |
10 msec |
|
ICD: |
10 msec |
WA64071, August 2009