Passive Flow Splitter (Tubing
Calculator)
|
|
|
|
|
Step 1
|
Input Flow rate, Split
Ratio and Solvent used
|
|
|
|
Step 2
|
Choose tubing ID for
Primary and Split from drop down (refer to table if needed)
|
|
|
Step 3
|
Choose length of tubing
desired for split flow path, it will calculate length of primary tubing
required
|
|
|
Observe time Offset,
Longer time difference is needed for fraction collection and a smaller time
offset.
|
|
|
|
is needed to align output
signals from two detectors. Adjust as necessary
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Input Total Flow Rate
|
|
|
mL/min
|
|
|
|
|
Input Split Ratio
desired
|
|
(eg for 50:1 split enter 50)
|
|
|
|
Solvent Name
|
|
|
|
|
|
Solvent viscosity (cp)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Primary Tube
|
Split Tube
|
|
|
Flow Rate (mL/min)
|
|
|
|
|
|
|
|
Tubing diameter (inch)
|
|
|
|
|
|
|
Tubing Length (feet)
|
|
|
|
|
|
|
|
Max Predicted pressure
drop (psi)
|
|
|
|
|
|
|
Time Offset (sec)
|
|
|
|
|
|
|
|
|
|
|
Tubing I.D.
(Inner Diameter)
|
O.D
|
|
|
|
|
Inch
|
µL/inch
|
mm
|
µL/cm
|
microns
|
|
|
|
|
|
0.093
|
109.57
|
2.362
|
43.83
|
2400
|
|
|
|
|
|
0.065
|
53.52
|
1.651
|
21.41
|
1650
|
1/8"
|
|
|
|
0.062
|
48.70
|
1.575
|
19.48
|
1600
|
(3.2mm)
|
|
|
0.058
|
42.62
|
1.473
|
17.05
|
1500
|
|
|
|
0.040
|
20.27
|
1.016
|
8.11
|
1000
|
|
|
|
0.030
|
11.40
|
0.762
|
4.56
|
760
|
|
|
|
0.020
|
5.07
|
0.508
|
2.03
|
500
|
|
|
|
0.013
|
2.14
|
0.320
|
0.80
|
320
|
|
|
|
0.010
|
1.27
|
0.254
|
0.51
|
250
|
1/16"
|
|
|
0.009
|
1.03
|
0.229
|
0.41
|
225
|
(1.6mm)
|
|
|
0.007
|
0.62
|
0.178
|
0.25
|
180
|
|
|
|
0.005
|
0.32
|
0.127
|
0.13
|
125
|
|
|
|
0.004
|
0.20
|
0.100
|
0.079
|
100
|
|
|
|
0.003
|
0.11
|
0.075
|
0.044
|
75
|
|
|
|
0.002
|
0.05
|
0.050
|
0.020
|
50
|
|
|
|
0.001
|
0.01
|
0.025
|
0.005
|
25
|
|
|
|