Analysis of Microcystins RR, LY, and YR in Bottled, Tap, and Surface Water Using ACQUITY UPLC Systems with 2D-LC Technology

Library Number:
APNT134826892
Part Number:
720005249EN
Author(s):
Claude Mallet
Source:
Waters
Content Type:
Application Notes
Content Subtype:
Application Notes
Related Products:
 
 
 
 

This application demonstrated the disruptive nature of the ACQUITY UPLC System with 2D-LC Technology with the Xevo TQD Mass Spectrometer. The application targeted the analysis of microcystin RR, LR, and YR in bottled, tap, and surface water. The limit of detection in this study was 50 ppt with a 10:1 enrichment from the extraction protocol (15 min total) and a 200:1 enrichment from the at-column dilution option, for a total of 2000:1. The recovery data for bottled, tap, and surface water samples using a microextraction protocol shows comparable results to applications with macroextraction protocols.


Title Format File Size
Download (720005249en) PDF 2453.96kB